Metrohm tiamo 2.4 Manual User Manual
Page 1617
■■■■■■■■■■■■■■■■■■■■■■
8 Manual control
tiamo 2.4
■■■■■■■■
1601
730, 774, 778, 789
Input range
0.01 - 160.00 mL/min
Titrino
Input range
0.01 - 150.00 mL/min
Filling rate
The buret will be filled at this rate after the dosing. The maximum filling
rate depends on the cylinder volume of the exchange unit or dosing
device used. If the entered filling rate is too high for the selected dosing
device, it will automatically be reduced during dosing to the largest possi-
ble value.
NOTE
The filling rate should be reduced for viscous liquids.
Selection
maximum
Default value
maximum
Titrando, 814, 815, 855
Input range
0.01 - 166.00 mL/min
730, 774, 778, 789
Input range
0.01 - 160.00 mL/min
Titrino
Input range
0.01 - 150.00 mL/min
Fill automatically
on | off (Default value: on)
This parameter is only visible for devices of the type Titrando, Dosing Inter-
face and USB Sample Processor.
If this option is activated, the buret is filled automatically after dosing.
During the filling procedure the volume display is reset to 0.0000 mL. If
this option is deactivated, the added volume is displayed cumulatively.
[Start]
Start fixed volume dosing for the selected dosing device. The dosed vol-
ume is displayed live.