Metrohm tiamo 2.4 Manual User Manual
Page 153
■■■■■■■■■■■■■■■■■■■■■■
3 Workplace
tiamo 2.4
■■■■■■■■
137
ID1…ID16
Sample identifications ID1…ID16. Any text may be entered into this field,
the checking for type and limit values does not take place until the start of
the determination. For sample identifications of the type Date/Time, the
entry must be made in the format YYYY-MM-DD or YYYY-MM-DD
hh:mm:ss. If text templates are defined for the sample identification,
then these can be selected (see Chapter 3.5.1, page 128).
Entry
100 characters
Selection
Selection of text templates
Sample size
Sample size (initial weight). Only numbers, decimal points, +, –, E and e
can be entered in this field. The checking for limit values does not take
place until the start of determination.
NOTE
If a value is entered or imported into this field, then in addition the time
of the data entry and the data source are stored manually, while the
'Name of the scale', the 'Name of the barcode reader' or the
'Name of the import file') are stored automatically at the same time,
placed in the determination and identified in the Information sub-
window as Data entry and Data source.
NOTE
If a negative value is entered in this field (e. g. during data import from
a scale in the event of reweighing), then this will be also be calculated
as a negative in the formula editor.
Entry
Number with a maximum of 10 digits
Default value
1.0
Unit
Sample size unit (initial weight). There are no restrictions on the text to be
entered in this field.
Entry
16 characters
Selection
g | mg | µg | mL | µL | pieces
Default value
g