Metrohm tiamo 2.4 Manual User Manual
Page 1043
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
1027
Moves the selected line down (changes the sequence).
[New]
Open the dialog window Fixed endpoint evaluation # to enter a new
fixed endpoint (see Chapter 5.6.4.10.6, page 1030).
[Properties]
Open the dialog window Fixed endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.4.10.6, page 1030).
[Delete]
Delete the selected line.
Minimum evaluation
Minimum evaluation
on | off (Default value: off)
If this option is activated, then the associated time and temperature for
the minimum measured value will be interpolated from the measuring
point list.
Threshold value
The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.
Input range
0.001 - 10.0 (mS/cm)/s
Default value
5.0 (mS/cm)/s
Maximum evaluation
Maximum evaluation
on | off (Default value: off)
If this option is activated, then the associated time and temperature for
the maximum measured value will be interpolated from the measuring
point list.
Threshold value
The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.
Input range
0.001 - 10.0 (mS/cm)/s
Default value