9 kft ipol - additional measured values – Metrohm tiamo 2.4 Manual User Manual
Page 820
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
804
■■■■■■■■
tiamo 2.4
Minimum evaluation
Minimum evaluation
on | off (Default value: off)
If this option is enabled, then the associated volume, time and tempera-
ture for the minimum measured value will be interpolated from the mea-
suring point list.
Threshold value
The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.
Input range
1.0 - 2000.0 mV/mL
Default value
25.0 mV/mL
Maximum evaluation
Maximum evaluation
on | off (Default value: off)
If this option is enabled, then the associated volume, time and tempera-
ture for the maximum measured value will be interpolated from the mea-
suring point list.
Threshold value
The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.
Input range
1.0 - 2000.0 mV/mL
Default value
25.0 mV/mL
5.6.3.5.2.9
KFT Ipol - Additional measured values
Tab: Method
▶ KFT Ipol ▶ Properties... ▶ Additional measured values
Command name
Name of the command.
Entry
25 characters
A maximum of 6 additional measured values can be defined on this tab.
These values can then be saved together in additional measured value col-
umns with the measured values present in the default settings.