4 tet - titration parameters – Metrohm tiamo 2.4 Manual User Manual
Page 938
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
922
■■■■■■■■
tiamo 2.4
Dosing rate
Rate (volume/time unit) at which the start volume is to be added. The
maximum dosing rate depends on the cylinder volume of the dosing unit
used.
Input range
0.100 - 166.000 mL/min
Selection
maximum
Default value
maximum
Pause
Pause
Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the dosing of a start volume. The pause follows at the
end of all the start conditions.
Input range
0 - 9999 s
Default value
0 s
5.6.3.9.4
TET - Titration parameters
Tab: Method
▶ TET ▶ Properties... ▶ Titration parameters
Command name
Name of the command.
Entry
25 characters
The parameters for the running of the titration are defined on this tab.
Dosing
Dosing rate
Rate (volume/time unit) at which dosing is to be carried out. The maxi-
mum dosing rate depends on the cylinder volume of the dosing unit used.
Input range
0.100 - 166.000 mL/min
Default value
2.000 mL/min
Selection
maximum
Filter
Filtering factor
Factor which determines the digital filtering of the measuring curve. The
higher this value, the more the noise is damped. The filtering has an influ-
ence on the shape of the peak and the EP shift.
Input range
1 - 200
Default value
50