4 dos ph - stop conditions, Stop conditions – Metrohm tiamo 2.4 Manual User Manual
Page 1195
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
1179
Input range
0 - 999999 s
Default value
0 s
Temperature
Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous,
then the temperature will be measured continuously. This value is used for
temperature correction during monitoring with pH measurements.
Titrando, 855
Input range
-20.0 - 150.0 °C
Default value
25.0 °C
Titrino
Input range
-170.0 - 500.0 °C
Default value
25.0 °C
Time interval measuring point
Time interval for entering a measuring point in the measuring point list.
Titrando, 855
Input range
0.1 - 999999 s
Default value
5.0 s
Titrino
Input range
1 - 999999 s
Default value
5 s
5.6.6.3.4.4
DOS pH - Stop conditions
Tab: Method
▶ DOS pH ▶ Properties... ▶ Stop conditions
Command name
Name of the command.
Entry
25 characters
The conditions for stopping the dosing are defined on this tab. If more
than one stop condition is in effect at the same time, then the criterion
which is fulfilled first will stop the dosing.
Stop volume
Stops when the given volume has been added after the start of the dos-
ing. The stop volume should be adapted to suit the sample weight or the
titration vessel size.