5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
646
■■■■■■■■
tiamo 2.4
EP recognition
Filter for the recognition of equivalence points:
Titrando, 855
Selection
first | greatest | last | ascending | descending
Default value
first
first
Only the first endpoint to be found will be recognized.
greatest
Only the endpoint with the greatest ERC value, i.e. the steepest jump,
will be recognized.
last
Only the last endpoint to be found will be recognized.
ascending
Only endpoints with a positive slope of the titration curve will be rec-
ognized.
descending
Only endpoints with a negative slope of the titration curve will be rec-
ognized.
5.6.3.3.4.11
MET Ipol - Fixed endpoint evaluation
Dialog window: Method
▶ MET Ipol ▶ Properties... ▶ Additional evalua-
tions
▶ Fixed endpoint evaluation ▶ [New]/[Properties] ▶ Fixed endpoint evalua-
tion #
Quantity
Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.
Selection
Measured value | Time | Volume
Default value
Measured value
Fixed value
Value of the fixed endpoint.
Measured value
Input range
-2000.0 - 2000.0 mV
Time
Input range
0.0 - 999999.9 s
Volume
Input range
0.00000 - 9999.99 mL