5 Method
tiamo 2.4
■■■■■■■■
987
5.6.4.6.4
MEAS Upol - Evaluations
Tab: Method
▶ MEAS Upol ▶ Properties... ▶ Evaluations
Command name
Name of the command.
Entry
25 characters
The following additional methods for evaluation of measurement curves
can be activated and defined on this tab.
Fixed endpoint evaluation
Fixed endpoint evaluation
on | off (Default value: off)
If this option is activated, then the associated values for the other mea-
sured quantities will be interpolated with a fix measured quantity (Mea-
sured value or Time) for the fixed endpoint from the measuring point
list.
Moves the selected line up (changes the sequence).
Moves the selected line down (changes the sequence).
[New]
Open the dialog window Fixed endpoint evaluation # to enter a new
fixed endpoint (see Chapter 5.6.4.6.6, page 991).
[Properties]
Open the dialog window Fixed endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.4.6.6, page 991).
[Delete]
Delete the selected line.
Minimum evaluation
Minimum evaluation
on | off (Default value: off)
If this option is activated, then the associated time and temperature for
the minimum measured value will be interpolated from the measuring
point list.