beautypg.com

Mpe and multipurpose testware (mpt), Mts mp and mts rpc – MTS Multipurpose Elite User Manual

Page 643

background image

Comparison of Trend Calculations to Other MTS
Applications

MPE and MultiPurpose TestWare (MPT)

Terminology Comparison

MPE and MPT Terminology Comparison

MPT Term

MPE Term

Smart Limits % Based on Data Value

Statistical Percentage

Smart Limits % Based on Full Range

Full-Scale Percentage

Eng. Units

Statistical Limits from Table

% Range

Range Percentage from Table

To emulate the default MPT table limits

1. On the workflow, click the Trend Monitoring activity.
2. On the Property panel, in the Limit Definition panel, set Upper Limits and Lower Limits to 100%.
3. For Calculation Method, select Statistical Limits from Table or Range Percentage from Table.

MTS MP and MTS RPC

The Statistic method in MTS RPC is the same as the Statistical Percentage method in MTS MP. The
Manual method in MTS RPC is the same as the Statistical Limits from Table method in MTS MPE.

Note:

MTS RPC does not define a Min value, so the range is not clearly defined.

RPC Full Scale method (using Min=-Max)

LowerLimit = Statistic – LowerPercentage * Range / 2

UpperLimit = Statistic + UpperPercentage * Range / 2

RPC Full Scale method (using Min=0.0)

LowerLimit = Statistic – LowerPercentage * Range

UpperLimit = Statistic + UpperPercentage * Range