Description

Control

Override

Caution

:

Station limits are disabled during the
override time.

Disabling station limits also disables
safety limits.

Always clear the actuator area before
applying hydraulics.

Temporarily override a test station interlock so that
you can clear the cause. Each time you click the
button, ten seconds is added to the length of time
the interlock is overridden, up to a total of 120
seconds. The digital readout to the right of the button
shows the duration and the countdown.

Power controls

Power Controls

Description

Control

Place the HSM (hydraulic service manifold) for the
test station and the HPU (hydraulic power unit) in
their zero-pressure states (red indicators lit).

Off

Place the HPU in the high-pressure state (green
indicator lit) and the HSM for the test station in the
low-pressure state (amber indicator lit).

Low

Place the HSM for the test station in the
high-pressure state (green indicator lit). You must
first click Low before you can click the High button.

High

MTS TestSuite | 611

Running Tests