Adding a digital input activity, Limit detection, Limit detection activity overview – MTS Multipurpose Elite User Manual
Page 401
Adding a Digital Input Activity
To add a Digital Input activity:
1. Add a Digital Input activity to the desired location in the test workflow.
2. From the Monitor list, select either the One Time or Continuous option.
3. Select a Trigger When option.
4. Select any controller action to take from the Action list.
5. Select any logging action to take from the Log list.
6. In the Digital Inputs panel, select the signal or signals you want to monitor and indicate the Options
(such as Low to High) for each digital input.
Limit Detection
Limit Detection Activity Overview
Limit Detection Activity Icon
The Limit Detection activity monitors signals and variables during a test run and compares their values
against defined upper and lower limits. Configure the Limit Detection activity to respond to a single limit
event or multiple limit events.
You can specify the action that the Limit Detection activity takes. For example, you can choose to indicate
that a limit is reached, or you can choose to stop the test and generate an interlock condition.
MTS TestSuite | 401
Test Activities