Control panel settings – MTS Multipurpose Elite User Manual
Page 614
Description
Control
Hold state, the button illuminates, and its colors
reverse.
Stop. Stop the test in place. The current cycle or
command activity is not completed. Click Run to
Stop
restart the test. You can unload a test that is stopped.
When the test is stopped, the button illuminates, and
its colors reverse.
Control Panel Settings
Access
Preferences menu > Configuration > Control Panel tab
Use the Control Panel tab to set up the operation of the Park and Unload buttons on the Application
control panel.
Note: When control is provided by a handset, the application controls will be locked and overlaid
by the handset exclusive control icon:
Control Panel Settings
Description
Item
This is the station (station.cfg) to which the MTS
TestSuite application is connected. If the application
Station
is not connected to a station, the controls in the
window are disabled.
Ramp Time and Commands set the length of time
an Enabled channel takes to achieve its programmed
End Level when you click Park or Unload.
Ramp Time
When you click Park or Unload, the application
determines the difference between the current signal
value and the target end level for each channel. It
then commands the channel to ramp to the target
end level at a rate based on the ramp time you enter.
This table includes the columns of parameters
necessary to set up the park and unload functions
Commands
for the available channels. Each column includes
ascending or descending sort options.
614 | MTS TestSuite
Running Tests