Create a signal meter, Create a signal meter for time, Signal scope – MTS Multipurpose Elite User Manual
Page 567: About signal scopes

Description
Item
Select the size of the label shown in the meter.
Label Font Size
Select the size of the value shown in the meter.
Value Font Size
Select a color (or transparent) of the foreground of the meter.
Foreground Color
Select a color (or transparent) of the background of the meter.
Background Color
Select Right, Center, or Left for the location of the text on the meter.
Text Alignment
Create a Signal Meter
To create a Signal Meter test-run display:
1. Click Monitor > Toolbox > Signal > Signal Meter. Drag the test-run display object to the Designer
tab.
2. (Optional) Change the default Display Name, or set a location with the Docking option.
3. (Required) Click the ellipsis in the Signal box and select a signal to monitor. The available signals
depend on the defined test resources.
4. Adjust the Value panel and Format panel properties as desired.
Create a Signal Meter for Time
This example creates a simple meter that shows total time elapsed during a test run. The meter continues
running even when a test is held.
1. Click Monitor > Toolbox > Signal > Signal Meter. Drag the test-run display object to the Designer
tab.
2. (Required) Click the ellipsis in the Signal box and select the Time signal. In the Value panel, the Meter
Type selection defaults to Timed, and the Unit to (sec).
Signal Scope
About Signal Scopes
The Signal Scope allows you to configure offline the scope properties during test runs. The signals that
are available in the scope properties come from the mapping in the test Resources. You can define the
scope configuration settings offline when the application is disconnected from the station. However, the
application must be connected to the controller when the test is run to show the data for the signals.
When a scope is used in a test-run display for a test:
•
The scope shows signals for the test when the test is run.
•
The scope updates only while the test is running and shows the last available signal status when the
test is stopped.
•
The scope configuration settings are saved with the test.
MTS TestSuite | 567
Test-Run Displays