beautypg.com

Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 985

background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

969

Threshold value

The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.

Range

0.1 ... 1000.0 mAU/s

Default value

20.0 mAU/s

Break-point evaluation

Break-point evaluation

on | off (Default value: off)

If this option is enabled, then sharp, almost 90° changes in directions in the
measurement curve will be evaluated.

EP criterion

Measure of the minimum sharpness of the break-point. The smaller the EP
criterion set, the more break-points will be found. As this is a relative value
related to the total measured value alteration, even small changes in the
measured value can be evaluated as a break-point for a small measured value
range.

Range

0.0 ... 1.0

Default value

0.3

Slope

Minimum difference between the slope before and after the break-point.
The smaller the difference, the more break-points will be found.

Range

0.0 ... 10.0

Default value

0.9

Smoothing factor

The higher the smoothing factor, the fewer endpoints will be found.

Range

2 ... 20

Default value

5

Window

A range (window) can be defined on the measured value axis or on the time
axis. The break-point evaluation will only be carried out in the defined win-
dow. Only the first break-point in the defined window will be recognized.

Selection

Measured value | Time | off

Default value

off

Lower limit

Value for the lower limit of the window.