Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 967
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.1 (ProcessLab)
■■■■■■■■
951
Parameters
The parameters for the command MEAS TC Cond are configured on the
following two tabs:
■
General/Hardware
Parameters for devices, sensors and stirrers.
■
Measuring parameters
Parameters for setting the measurement procedure.
■
Additional measured values
Defining further measured values which can be saved as additional col-
umns in the measuring point list.
Command variables
The following command variables are generated in the method run by the
command MEAS Cond and can be used in formulas under the designation
'Command name.Variable designation':
Identification
Description
.BP{x}.MEA
Measured value for the break-point (1…9) in the unit
of the measured value
.BP{x}.TEM
Temperature for the break-point × (1…9) in °C
.BP{x}.TIM
Time for the break-point × (1…9) in s
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has not ever been started
.DBL
Total duration for the processing of the command in s
.EME
Final measured value (measured value after processing
of the command) in the unit of the measured value
.ENP
Electrode zero point of the sensor used in the command
.ETE
End temperature (temperature following processing of
the command) in °C
.FIN
Command status; 1 = Command has been ended
at least once, 0, invalid (variable not available) =
Command has never been ended
.FP{x}.MEA
Measured value for the fix endpoint × (1…9) in the unit
of the measured value
.FP{x}.TEM
Temperature for the fix endpoint × (1…9) in °C