4 meas u - evaluations, Evaluations – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 897
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.1 (ProcessLab)
■■■■■■■■
881
off
No stop.
Temperature
Temperature
Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous, then
the temperature will be measured continuously.
Titrando, 855, 867
Range
-20.0 ... 150 °C
Default value
25.0 °C
Titrino
Range
-170.0 ... 500 °C
Default value
25.0 °C
5.6.4.3.4
MEAS U - Evaluations
Tab: Method
▶ MEAS U ▶ Properties... ▶ Evaluations
Command name
Name of the command.
Input
25 characters
The following additional methods for evaluation of measurement curves can
be activated and defined on this tab.
Fix endpoint evaluation
Fix endpoint evaluation
on | off (Default value: off)
If this option is activated, then the associated values for the other measured
quantities will be interpolated with a fix measured quantity (Measured
value or Time) for the fix endpoint from the measuring point list.
Moves the selected line up (changes the sequence).
Moves the selected line down (changes the sequence).
[New]
Opens the dialog window Fix endpoint evaluation # to enter a new fixed
endpoint (see Chapter 5.6.4.3.6, page 885).