Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 851
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.1 (ProcessLab)
■■■■■■■■
835
Dosing device
Selection of the number of the dosing device (dosing or exchange unit) to
be used for the dosing. All the dosing device connections which are possible
with the selected device type are displayed.
Titrando
Selection
1 | 2 | 3 | 4
Default value
2
855
Selection
1 | 2 | 3
Default value
2
Solution
Input of a solution name or selection of a solution from the solutions defined
in the solution table. If intelligent exchange or dosing units are used, then
a check will be made in the method run to see whether the correct solution
has been set on the selected dosing device and whether the dosing device
type is correct. With non-intelligent exchange or dosing units, only the cyl-
inder volume is checked.
Titrando, 855
Selection
'Solution name' | not defined
Default value
not defined
not defined
No tests will be carried out.
Filling rate
Speed with which the dosing cylinder of the second dosing device is to be
refilled. The maximum filling rate depends on the cylinder volume of the
exchange unit or dosing device used. If the entered filling rate is too high
for the selected dosing device then during dosing it will automatically be
reduced to the largest possible value ((see Chapter 5.6.3.8.5, page 803)).
Titrando, 855
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
Sensor
Measuring input
Selection of the measuring input to which the sensor is connected.