5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

932

■■■■■■■■

tiamo 2.1 (ProcessLab)

Device type

Display or selection of the device type. If a device is selected under Device
name
, then this Device type field can no longer be edited, and the device
type belonging to the selected device is displayed instead

If the option not defined is selected as Device name, any device types or
device groups which are able to execute the command can be selected,
independently of the devices in the device table.

Selection

Device types | Titrando

Default value

Titrando

Sensor

Measuring input

Selection of the measuring input to which the sensor is connected.

Titrando, 867
Selection

1 | 2

Default value

1

855
Selection

1

Default value

1

Sensor

Selection of an ISE sensor from the ISE sensors available in the sensor table.

Selection

'Sensor name' | ISE electrode

Default value

ISE electrode

Temperature measurement

Type of temperature measurement.

Titrando, 855
Selection

continuous | automatic | off

Default value

automatic

continuous
A temperature sensor must be connected. The temperature will then be
measured continuously.
automatic
If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise the temperature entered manually under
Temperature on the Titration parameters tab will used.