8 det upol - additional measured values – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 548
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
532
■■■■■■■■
tiamo 2.1 (ProcessLab)
Upper limit
Value for the upper limit of the Gran evaluation range.
Range
-200.0 ... 200.0 µA
Default value
200.0 µA
5.6.3.2.5.8
DET Upol - Additional measured values
Tab: Method
▶ DET Upol ▶ Properties... ▶ Additional measured values
Command name
Name of the command.
Input
25 characters
A maximum of 6 additional measured values can be defined on this tab
which can then be saved together in additional measured value columns
with the measured values present in the default settings:
Additional calculated measured values
Additional calculated measured values
on | off (Default value: off)
If this option is enabled, then a maximum of 3 new measured values can be
calculated from existing measured values and saved, which can also be pre-
sented as curves under the designation Calculated 1…3 and used in
formulas as variables 'Command name.CA1…3'.
[New]
Opens the dialog window Calc. measured value # in which the parame-
ters for the additional measured value can be entered (see Chapter
5.6.3.10.1, page 858).
[Properties]
Opens the dialog window Calc. measured value # in which the parame-
ters for the additional measured value can be edited (see Chapter
5.6.3.10.1, page 858).
[Delete]
Deletes the calculated measured value selected in the table.
Additional external measured values
Additional external measured values
on | off (Default value: off)
If this option is enabled, then a maximum of 3 new measured values can be
taken over and saved from existing measuring commands (i.e. MEAS) which