5 cal cond - data conductivity standard, 9 cal spec, 1 cal spec - overview – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 1047: Cal spec
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.1 (ProcessLab)
■■■■■■■■
1031
Name standard
List box with conductivity standards. The data of one of these standards is
used to calculate the temperature coefficient for the specified reference
temperature.
Selection
'Conductivity standard'
5.6.5.8.5
CAL Cond - Data Conductivity standard
Dialog window: Method
▶ CAL Cond ▶ Data Conductivity standard
Current temperature
Display of the currently measured temperature of the conductivity standard.
Data transmission from standard
Input of the data of the conductivity standard, which should be as close as
possible to the Current temperature. This data is as a rule contained in
the Data Sheet/Certificate for the conductivity standard.
Reference temperature
Input of a temperature for which a calibration point of the conductivity
standard is known. It should be as close as possible to the temperature at
which measurements are carried out.
Range
-20.0 ... 150.0°C
Default value
20.0
Conductivity
Input of the conductivity of the standard at the reference temperature.
Range
0.00000 ... 2000.00 mS/cm
Default value
1.0 mS/cm
Temperature coefficient
Temperature coefficient of the standard with the reference temperature.
Range
0.00 ... 9.99 %/°C
Default value
1.00 %/°C
5.6.5.9
CAL Spec
5.6.5.9.1
CAL Spec - Overview
Dialog window: Method
▶ CAL Spec ▶ Properties... ▶ CAL Spec - 'Command name'
Command for wavelength calibration.
Devices
This command can be executed with the following devices:
Spectrometer: Avantes