6 cal loop conc, 1 cal loop conc - overview, Cal loop conc – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 1033
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.1 (ProcessLab)
■■■■■■■■
1017
Range
-20.0 ... 150.0 °C
Default value
25.0 °C
5.6.5.6
CAL LOOP Conc
5.6.5.6.1
CAL LOOP Conc - Overview
Dialog window: Method
▶ CAL LOOP Conc ▶ Properties.. ▶ CAL LOOP Conc - 'Com-
mand name'
Loop command for the calibration of ion-selective electrodes, which is
used together with the command CAL MEAS Conc.
Appearance
The command has the following appearance:
Parameters
The parameters for the command CAL LOOP Conc are configured in the
following dialog window:
■
Command variables
The following command variables are generated in the method run by the
command CAL LOOP Conc and can be used in formulas under the desig-
nation 'Command name.Variable designation':
Identifi-
cation
Description
.BLV
Blank value calculated from the calibration
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command has
not ever been started
.ENP
Electrode zero point calculated from the calibration
.FIN
Command status; 1 = Command has been ended at
least once, 0, invalid (variable not available) = Command
has never been ended
.LCO
Loop counter = current quantity of completed runs, both of
Repeat loops and of While loops
.LST
Start time of the loop command (Date/Time)
.SLO
Electrode slope calculated from the calibration