3 curves - monitoring report, Curves - monitoring report – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 371
■■■■■■■■■■■■■■■■■■■■■■
4 Database
tiamo 2.1 (ProcessLab)
■■■■■■■■
355
The display of the measuring points in the table can be defined per window
and per command type in the Properties - Curve #
4.8.3
Curves - Monitoring report
Dialog window: Database
▶ Curves
If the method contains commands of the type STAT or DOS, then with the
context-sensitive menu item Monitoring report... in the curve window the
dialog window Monitoring report - 'Command name' can be opened.
In this window information on limit violations occurred are displayed for
each monitoring switched on
The following columns are displayed:
Time [s]
Time in s until limit violation.
Limit
Selection
lower | upper | ok
lower
Lower limit value exceeded.
upper
Upper limit value exceeded.
ok
Measured value is inside limits again.
[pH]/[U]
Measured value limit.
Meas.v. [pH]/Meas.val. [U]
Measured value at limit violation time.
Volume [mL]
Dosing volume at limit violation time.
Temp. [°C]
Temperature at limit violation time.
Action
Selection
Stop determination | Stop command (only
Titrando) | Wait for [Continue] | Wait for limit
ok | none
Stop determination
The running STAT command was quit, then the determination was
stopped.