4 met ipol - titration parameters, Titration parameters – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 615
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.1 (ProcessLab)
■■■■■■■■
599
Selection
maximum
Start slope
Note
Is displayed only for Titrando and 855.
Start slope
When the start slope is reached the dosing of the start volume is stopped
and the titration is started. If the start slope is achieved by the dosing of a
start volume then the titration starts directly.
Range
0 ... 999 mV/mL
Selection
off
Default value
off
Dosing rate
Speed at which the start volume is added until the start slope is reached.
The maximum dosing rate depends on the cylinder volume of the exchange
unit or dosing unit used.
Range
0.01 ... 166.00 mL/min
Default value
5.00 mL/min
Selection
maximum
Pause
Pause
Waiting time, e.g. for the electrode to settle down after the start or a reac-
tion time after the addition of the start volume. The pause follows at the
end of all the start conditions.
Range
0 ... 999999 s
Default value
0 s
5.6.3.3.4.4
MET Ipol - Titration parameters
Tab: Method
▶ MET Ipol ▶ Properties... ▶ Titration parameters
Command name
Name of the command.
Input
25 characters
Parameters defining the run of the titration.