Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 1123
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.1 (ProcessLab)
■■■■■■■■
1107
Device
Device name
Selection of a device from those available in the device table. The only devi-
ces which shall be offered are those with which the command can run.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start of
the method.
Device type
Display or selection of the device type. If a device is selected under Device
name, then this Device type field can no longer be edited, and the device
type belonging to the selected device is displayed instead.
If the option not defined is selected as Device name, any device types or
device groups which are able to execute the command can be selected,
independently of the devices in the device table.
Selection
774 Oven Sample Processor | 874 Oven Sample
Processor
Default value
874 Oven Sample Processor
Heater
Target temperature
Temperature to which the oven has to be heated. With init.temp. the oven
heating is switched on at the same time as the device and the oven is heated
up to this temperature. The initial temperature must be defined on the device
itself and is displayed only in the device configuration.
Range
50 ... 250 °C
Selection
Initial temp.
Default value
Initial temp.
Note
It is advisable to define an initial temperature if you work always at the
same temperature.
Heating time
The target temperature has to be reached within this time. The heating rate
can be modified with the heating time in order to program temperature
ramps or entire temperature profiles (with several HEATER commands).