5 Method
tiamo 2.1 (ProcessLab)
■■■■■■■■
437
Sensor
Selection of a sensor of type pH electrode from the list of sensors available
in the sensor table. The calibration data for the sensor will be adopted for
the determination.
Selection
Sensor name | pH electrode
Default value
pH electrode
Temperature measurement
Type of temperature measurement.
Titrando, 855
Selection
continuous | automatic | off
Default value
automatic
continuous
A temperature sensor must be connected. The temperature will then be
measured continuously.
automatic
If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise the temperature entered manually under
Temperature on the Titration parameters tab will used.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Titration parameters tab will be
used.
Stirrer
Stirrer
Selection of the stirrer.
Titrando, 855
Selection
1 | 2 | 3 | 4 | off
off
No stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Titrando, 855
Range
-15 ... 15
Default value
8