3 meas t/flow - measuring parameters, 4 meas t/flow - evaluations, Measuring parameters – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 943: Evaluation
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.1 (ProcessLab)
■■■■■■■■
927
5.6.4.7.3
MEAS T/Flow - Measuring parameters
Tab: Method
▶ MEAS T/Flow ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Input
25 characters
The parameters for the measuring procedure can be defined on this tab.
Measuring time
Maximum period of time for measurement.
Range
0 ... 999999 s
Default value
600 s
Measuring interval
Time interval for entering a measuring point in the measuring point list.
Range
1 ... 999999.0 s
Default value
10 s
Stop measured value Temp.
Stops when the preset measured value has been reached since the start of
the measurement. off means no stop.
Range
50.0 ... 250.0 °C
Selection
off
Default value
off
5.6.4.7.4
MEAS T/Flow - Evaluations
Tab: Method
▶ MEAS T/Flow ▶ Properties... ▶ Evaluations
Command name
Name of the command.
Input
25 characters
The following additional methods for evaluation of measurement curves can
be activated and defined on this tab.
Minimum evaluation
Minimum evaluation
on | off (Default value: off)
If this option is enabled, then the associated time for the minimum measured
value will be interpolated from the measuring point list.