Identification
Description
.COK
Command status; 1 = Condition requirement ful-
filled, 0 = Condition requirement not fulfilled
.CONC
Concentration of the solution used for the command
.CYL
Cylinder volume of the exchange or dosing unit used
for the command
.DBL
Total duration for the processing of the command in s
.DRI
Current and/or last drift for drift correction in µL/min
.DSC
Time for processing all start conditions s
.DTI
Time for drift correction (time from the start of the
titration to the end of the command) in s
.EME
End measured value (measured value after processing
of the command) in the unit of the measured value
.EP{x}.MEA
Measured value for the endpoint × (1…9) in the unit
of the measured value
.EP{x}.MEP
Quantity of endpoints in the window × (1…9); 1 = 1
endpoint, 2 = 2 or more endpoints, 3 = EP cor-
rected with Autodrift, 4 = EP corrected with
manual drift
.EP{x}.TEM
Temperature for the endpoint × (1…9) in °C
.EP{x}.TIM
Time in s until the endpoint × (1…9) is reached
.EP{x}.VOL
Volume for the endpoint × (1…9) in mL
.ETE
End temperature (temperature following processing of
the command) in °C
.EVT
End volume (total dosed volume at the end of the com-
mand) in mL
.FIN
Command status; 1 = Command has ended at least
once, 0, invalid (variable not available) = Command
has never ended
.FP{x}.MEA
Measured value for the fixed endpoint × (1…9) in mV
.FP{x}.TEM
Temperature for the fixed endpoint × (1…9) in °C
.FP{x}.TIM
Time in s until the fixed endpoint × (1…9) is reached
.FP{x}.VOL
Volume for the fixed endpoint × (1…9) in mL