5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
790
■■■■■■■■
tiamo 2.1 (ProcessLab)
continuous
A temperature sensor must be connected. The temperature will then be
measured continuously.
automatic
If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise the temperature entered manually under
Temperature on the Titration parameters tab will used.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Titration parameters tab will be
used.
Cell
Generator type
Selection of the generator type.
Selection
without diaphragm | with diaphragm
Default value
with diaphragm
Generator current
Polarization current at the generator electrode.
Selection
100 mA | 200 mA | 400 mA | auto
Default value
auto
auto
Means that the current is automatically adapted to the conductivity of
the reagent and that in the region of the endpoint the current will be
controlled at smaller values
Stirrer
Stirrer
Selection of the stirrer.
Selection
1 | 2 | 3 | 4 | off
Default value
1
off
No stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Range
-15 ... 15
Default value
8