Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 755
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.1 (ProcessLab)
■■■■■■■■
739
Range
0.1 ... 9.9 µL/min
Default value
minimum (slow), minimum (optimal), 5.0
(fast) µL/min
Stop criterion
Stop criterion
The titration us stopped after the last dosing once the endpoint has been
reached and the stop criterion has been fulfilled. The titration can be stopped
when a particular drift has been achieved or after a preset time.
Titrino
Selection
drift | time
Default value
drift
Titrando
Selection
drift | rel. drift | off | time
Default value
drift
drift
The titration is stopped when the stop drift has been reached.
rel. drift
The titration is stopped when the relative stop drift (drift value at time of
start + relative stop drift) has been reached.
off
The titration will not be stopped.
time
The endpoint must be exceeded for a certain time: the Delay time. The
same Delay time with different, very small volume increments (depending
on the volume of the exchange or dosing unit) means different switch-
off points.
Note
The Stop conditions will always stop the titration even if the stop criterion
has not been fulfilled or has been switched off.
Stop drift
Titration is stopped when the endpoint and the stop drift have been reached.
This parameter is only shown for Stop criterion = drift.
Range
1 ... 999 µL/min
Default value
20 µL/min