Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 838
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
822
■■■■■■■■
tiamo 2.1 (ProcessLab)
Wait for [Continue]
Reagent dosing in the current STAT command is interrupted and a mes-
sage appears. As soon as the monitored measured value is again within
the limits (including hysteresis) reagent dosing can be resumed by press-
ing the [Continue] button in this message box.
Wait for limit ok
Reagent dosing in the current STAT command is interrupted. As soon
as the monitored measured value is again within the limits (including
hysteresis), reagent dosing will be resumed automatically.
none
No action will be taken if limits are exceeded.
Monitoring Dosing rate
Monitoring Dosing rate
on | off (Default value: off)
If this option is activated, then the dosing rate will be monitored and any
exceeded limits will be entered in the list of measuring points. This moni-
toring is not started until 10 s after the start of the titration.
Lower limit
Lower limit of the dosing rate. When the dosing rate falls below this value,
the action Dosing rate lower limit exceeded will be triggered.
Titrando, 855
Range
0.00 ... 166.00 mL/min
Default value
0.00 mL/min
Titrino
Range
0.00 ... 150.00 mL/min
Default value
0.00 mL/min
Lower hysteresis
The lower hysteresis represents a tolerance range for the lower limit of the
dosing rate. When the dosing rate exceeds the lower limit by this hysteresis
value, the action Dosing rate lower limit ok is triggered.
Titrando, 855
Range
0.00 ... 166.00 mL/min
Default value
0.20 mL/min
Upper limit
Upper limit of the dosing rate. If the dosing rate exceeds this limit, then the
event Dosing rate upper limit exceeded will be triggered.