5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

992

■■■■■■■■

tiamo 2.1 (ProcessLab)

The parameters for the measuring procedure can be defined on this tab.

Measurement with drift control
The measurement is stopped as soon as the defined Signal drift or the
Stop measured value is reached or a defined Waiting time has elapsed.

Measured value drift

The measured value is only accepted if the signal drift defined here has been
fallen short of.

Range

0.1 ... 999.0 mV/min

Default value

10.0 mV/min

Min. waiting time

The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The drift continues
to be checked while the waiting time is elapsing.

Range

0 ... 999999 s

Default value

0 s

Max. waiting time

If the signal drift has not yet been reached, then the measured value will be
accepted when the maximum waiting time has elapsed. If the waiting time
has not been newly entered then a waiting time that is suitable for the drift
will be calculated automatically according to the following equation:

Range

0 ... 999999 s

Default value

52 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

Range

0.1 ... 999999.0 s

Default value

2.0 s

Temperature

Temperature

Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous, then
the temperature will be measured continuously.

Range

-20.0 ... 150°C

Default value

25.0