5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1110
■■■■■■■■
tiamo 2.1 (ProcessLab)
Device type
Display or selection of the device type. If a device is selected under Device
name, then this Device type field can no longer be edited, and the device
type belonging to the selected device is displayed instead.
If the option not defined is selected as Device name, any device types or
device groups which are able to execute the command can be selected,
independently of the devices in the device table.
Selection
774 Oven Sample Processor | 874 Oven Sample
Processor
Default value
874 Oven Sample Processor
Gas
Gas type
Gas type of the carrier gas.
Selection
Air | Nitrogen | other gas
Default value
Air
Gas flow factor
Factor for the correct determination of the gas flow. This parameter is only
visible for Gas type = other gas. The values for frequently used carrier gases
are listed in the device manuals.
Range
0.001 ... 9.999
Default value
1.000
Gas flow
Inlet
Selection of the inlet of the carrier gas
Selection
Pump | Valve
Default value
Pump
Pump
Use of ambient air as carrier gas.
Valve
Use of gas out of a compressed gas cylinder.
Flow rate
874
Range
10 ... 150 mL/min
Default value
50 mL/min