5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
774
■■■■■■■■
tiamo 2.1 (ProcessLab)
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Titration parameters tab will be
used.
Cell
Generator type
Selection of the generator type.
Selection
without diaphragm | with diaphragm
Default value
without diaphragm
Generator current
Polarization current at the generator electrode.
für Generator type = without diaphragm
Selection
100 mA | 200 mA | 400 mA | auto
Default value
400 mA
auto
Means that the current is automatically adapted to the conductivity of
the reagent and that in the region of the endpoint the current will be
controlled at smaller values
für Generator type = with diaphragm
Selection
100 mA | 200 mA | 400 mA | auto
Default value
auto
auto
Means that the current is automatically adapted to the conductivity of
the reagent and that in the region of the endpoint the current will be
controlled at smaller values
Note
For generator electrodes of the type with diaphragm, it is recommen-
ded to set the Generator current to auto.
For generator electrodes without diaphragm it is recommended to set
the Generator current to 400 mA.
Stirrer
Stirrer
Selection of the stirrer.