8 properties - glp – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 1467
■■■■■■■■■■■■■■■■■■■■■■
7 Devices
tiamo 2.1 (ProcessLab)
■■■■■■■■
1451
Default value
Display of the default value the outputs takes directly after starting and exit-
ing the program. The default value is also set after an error or when canceling
the method.
Lastly used value
on | off (Default value: off)
If this option is enabled the lastly set value is used as default value on the
analog output.
Enter the default value.
Range
-214783647 ... 214783647
Default value
0
7.19.8
Properties - GLP
Tab: Configuration
▶ Device ▶ [Edit] ▶ Properties... ▶ 'Properties - 'Device type' -
'Device name'
▶ GLP
GLP test date
Date of the last GLP test. This date can be selected by clicking on
in the
dialog window Select date (see Chapter 2.5.1, page 79).
Comment GLP test
Comment on GLP test.
Input
1000 characters
Monitoring of GLP validity
Monitoring of GLP validity
on | off (Default value: off)
If this option is switched on then the time interval for the GLP test will be
monitored.
Monitoring can only be switched on when a date has been entered in the
field GLP test date.
GLP test interval
Time interval to next GLP test. If a value is entered here then the date in the
field Next GLP test will be adapted automatically.
Only editable when the monitoring is activated.
Range
1 ... 999 Days
Default value
999 Days