Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 555
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.1 (ProcessLab)
■■■■■■■■
539
Identification
Description
.DSC
Time for processing all start conditions s
.EME
End measured value (measured value after processing
of the command) in the unit of the measured value
.ENP
Electrode zero point of the sensor used for the com-
mand (dimensionless)
.EP{x}.DME
Measured value differential for the endpoint × (1…9)
.EP{x}.MEA
Measured value for the endpoint × (1…9) in the unit
of the measured value
.EP{x}.MEP
Quantity of endpoints in the window × (1…9); 1 = 1
endpoint, 2 = 2 or more endpoints, 3 = EP cor-
rected with Autodrift, 4 = EP corrected with
manual drift
.EP{x}.TEM
Temperature for the endpoint × (1…9) in °C
.EP{x}.TIM
Time in s until the endpoint × (1…9) is reached
.EP{x}.VOL
Volume for the endpoint × (1…9) in mL
.ETE
End temperature (temperature following processing of
the command) in °C
.EVT
End volume (total dosed volume at the end of the com-
mand) in mL
.FIN
Command status; 1 = Command has ended at least
once, 0, invalid (variable not available) = Command
has never ended
.FP{x}.DME
Measured value differential for the fixed endpoint ×
(1…9)
.FP{x}.MEA
Measured value for the fixed endpoint × (1…9) in mV
.FP{x}.TEM
Temperature for the fixed endpoint × (1…9) in °C
.FP{x}.TIM
Time in s until the fixed endpoint × (1…9) is reached
.FP{x}.VOL
Volume for the fixed endpoint × (1…9) in mL
.GP.VOL
Volume for the Gran endpoint in mL
.GP.MEA
Measured value for the Gran endpoint in the unit of the
measured value
.GP.TEM
Temperature for the Gran endpoint in °C
.GP.TIM
Time in s until the Gran endpoint is reached