5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1038
■■■■■■■■
tiamo 2.1 (ProcessLab)
Titrando, 846, 856, 867
Selection
1 | 2 | 3 | 4
Default value
1
778, 789, 814, 815, 855, 864, 874
Selection
1 | 2 | 3
Default value
1
730, 774
Selection
1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12
Default value
1
Titrino
Selection
internal D0
736, 751, 758, 799
Selection
internal D0 | external D1 | external D2
Default value
internal D0
Solution
Input of a solution name or selection of a solution from the solutions defined
in the solution table. If intelligent exchange or dosing units are used, then
a check will be made in the method sequence to see whether the correct
solution has been set on the selected dosing device and whether the dosing
device type is correct. With non-intelligent exchange or dosing units, only
the cylinder volume is checked. At the start of the command, a check is made
of the working life, the validity of the titer and the GLP test interval for the
selected solution.
Selection
'Solution name' | not defined
Default value
not defined
not defined
No tests will be carried out.
Tandem dosing
on | off (Default value: off)
If this option is enabled, the dosing is carried out without interruption with
a combination of two dosing devices so that the second dosing device is
dosing while the first one is being filled and vice-versa. This parameter is
displayed only for Titrandos, 846, 856, 867, 814, 815, 855, 864 and 874.
Dosing device
Selection of the second dosing device (exchange or dosing unit) to be used
for the dosing if the first dosing device is not available.