6.6 Subwindow Titrants/Solutions
■■■■■■■■■■■■■■■■■■■■■■
1262
■■■■■■■■
tiamo 2.1 (ProcessLab)
Rel. standard dev. (s rel)
Relative standard deviation for the titer determination in %.
Monitoring titer validity
on | off (Default value: off)
If this option is switched on then the validity of the titer will be monitored.
Validity
Validity of the titer in days. If a value is entered here then the date for the
Next titer determination will be automatically adapted.
Range
0 ... 999 Days
Default value
999 Days
Next titer determination
Date on which the next titer determination is to be carried out. The date can
be selected by pressing
in the dialog window Select date (see Chapter
2.5.1, page 79). After the date has been entered, the Validity field will be
adapted automatically.
Set value
Date Titer det.+ 999 days
Message
The message options can only be edited if monitoring is switched on.
Message by E-mail
on | off (Default value: off)
The message is additionally sent to the E-mail address defined under [E-
mail] if this option is activated. The message is sent in text format.
[E-mail]
[E-Mail] opens the Send E-mail window (see Chapter 2.6.1, page 81).
Acoustic signal
on | off (Default value: off)
If this option is enabled, an acoustic signal will be emitted additionally to the
message defined above
Action
The actions can only be edited if monitoring is switched on.
If during titer monitoring it is found that the validity period of the titer has
expired then one of the following actions will be triggered automatically at
Start test.