3 meas opt - measuring parameters – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 982
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
966
■■■■■■■■
tiamo 2.1 (ProcessLab)
Device
Device name
Selection of a device from those available in the device table. The only devi-
ces which shall be offered are those with which the command can run.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start of
the method.
Device type
Display or selection of the device type. If a device is selected under Device
name, then this Device type field can no longer be edited, and the device
type belonging to the selected device is displayed instead
If the option not defined is selected as Device name, any device types or
device groups which are able to execute the command can be selected,
independently of the devices in the device table.
Selection
Avantes Spectrometer
Default value
Avantes Spectrometer
5.6.4.13.3
MEAS Opt - Measuring parameters
Tab: Method
▶ MEAS Opt ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Input
25 characters
The parameters for the measuring procedure can be defined on this tab.
Wavelength
Wavelength at which the measured quantity (absorbance) is recorded
Range
200.0 ... 1100.0 nm
Default value
400.0 nm
Integration time
Integration time for absorption measurement.
Range
0.01 ... 600000 ms
Default value
6 ms
Averaged spectra
The number of spectra that are recorded and averaged.