10 kft ipol - fix endpoint evaluation, 3 kft upol, 1 kft upol - overview – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 764: Kft upol
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
748
■■■■■■■■
tiamo 2.1 (ProcessLab)
5.6.3.5.2.10
KFT Ipol - Fix endpoint evaluation
Dialog window: Method
▶ KFT Ipol ▶ Properties... ▶ Additional evaluations ▶ Fix
endpoint evaluation
▶ [New]/[Property] ▶ Fix endpoint evaluation #
Quantity
Selection of the fixed measured quantity to which the associated values for
the other quantities are to be interpolated from the list of measured points.
Selection
Measured value | Time | Volume
Default value
Measured value
Fix value
Value of the fix endpoint.
Measured value
Range
-2000.0 ... 2000.0 mV
Time
Range
0.0 ... 999999.9 s
Volume
Range
0.00000 ... 9999.99 mL
5.6.3.5.3
KFT Upol
5.6.3.5.3.1
KFT Upol - Overview
Dialog window: Method
▶ KFT Upol ▶ Properties... ▶ KFT Upol - 'Command name'
Command for Karl Fischer Titrations with amperometric measurement
(selectable polarization voltage).
Devices
This command can be executed with the following devices.
Titrando: 835, 836, 841, 852, 857, 890, 901, 906, 907
Titrino: 720, 736, 751, 758, 784, 795, 799
Appearance
The command has the following appearance:
without conditioning
with conditioning