3 meas upol - measuring parameters, Measuring parameters – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 919
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.1 (ProcessLab)
■■■■■■■■
903
automatic
If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise, the temperature entered manually under
Temperature on the Measuring parameters tab will be used.
off
The temperature will not be measured. The temperature manually
entered under Temperature on the Measuring parameters tab will
be used.
Stirrer
Stirrer
Selection of the stirrer.
Titrando, 855, 867
Selection
1 | 2 | 3 | 4 | off
off
No stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Titrando, 855, 867
Range
-15 ... 15
Default value
8
Switch off automatically
on | off (Default value: on)
If this option is enabled, the stirrer will be switched off automatically when
the command has finished. This parameter is displayed only for Titrandos,
855 and 867.
5.6.4.5.3
MEAS Upol - Measuring parameters
Tab: Method
▶ MEAS Upol ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Input
25 characters
The parameters for the measuring procedure can be defined on this tab.
Measurement
The two measuring modes Measurement with drift control or Meas-
urement without drift control can be selected.