6 Configuration
tiamo 2.1 (ProcessLab)
■■■■■■■■
1231
6.3.4.2
Options - General
Tab: Configuration
▶ Tools ▶ Options ▶ Options ▶ General
Dialog language
Dialog language
Selection of the dialog language.
Selection
German | English | additional languages
(dependent on installed language patches)
Default value
English
Note
For the altered setting to become effective the program must be restarted.
Emergency stop button
If this option is enabled then the emergency stop button will be shown in
all program parts.
Selection
on | off
Default value
off
This button can be moved to any position with the left-hand mouse button
pressed down and is always shown right up at the top on the screen, both
in the program window and outside it on the Windows desktop. A mouse
click on this button immediately stops all running determinations on all acti-
vated workplaces. When this happens, all devices are stopped (including
pumps; any device commands that may have been started will be carried
out to their ends, e.g. PREP with Dosino) and the exit track will be started.
The emergency stop is effective even if no user is logged in, e.g. when the
user has been logged out automatically.
The emergency stop has no effect on running actions in the manual control,
each of which must therefore be stopped with
6.3.4.3
Options - Save
Tab: Configuration
▶ Tools ▶ Options ▶ Options ▶ Save
Save on closing
Here you can define the settings to be saved when the program is exited. If
the option is switched on the current view with its settings will be saved
automatically when the program is exited. If the option is switched off