5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

1133

Titrando, 856, 867, 814, 815, 855, 864, 874
Selection

1 | 2 | 3 | 4

Default value

1

Scan lines

Input signal

Input of the binary pattern for the expected input signal or selecting a pre-
defined signal pattern.

It is possible to enter the following characters:

0 = line inactive

1 = line active

* = any line status

Selection

Bit-pattern with 8 characters (0, 1, *) | ********
| Signal pattern

Default value

********

The bits are numbered from right to left:

Input 7 6 5 4 3 2 1 0

Bit

7 6 5 4 3 2 1 0

Example:

*******1 expects an active input line 0 (1 = set). For example, this line is set
by a Titrino after a titration has been completed and the Titrino can accept
a start signal again.

Note

Input lines that are of no interest or for which no defined condition can
be predicted should also be masked with an asterisk *.

Max. waiting time

on | off (Default value: off)

If this option is activated, then a maximum waiting time can be entered for
the input signal.

Range

0 ... 9999.9

Default value

10

Selection

min | s

Default value

s