5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
518
■■■■■■■■
tiamo 2.1 (ProcessLab)
Titrando, 855
Range
0.01 ... 99.90 µA/min
Default value
20.00 (slow), 50.00 (optimal), 80.00 (fast) µA/
min
Selection
off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed.
Titrino
Range
0.05 ... 99.90 µA/min
Default value
20.00 (slow), 50.00 (optimal), 80.00 (fast) µA/
min
Selection
off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed.
Min. waiting time
The minimum waiting period is only important for drift-controlled measure-
ments. The measured value is not accepted until the minimum waiting time
has elapsed, even if the signal drift has already been reached. The drift con-
tinues to be checked while the waiting time is elapsing.
Range
0 ... 999999 s
Default value
0 s
Max. waiting time
If signal drift has been switched off or has not yet been reached, then the
measured value will be accepted when the maximum waiting time has
elapsed.
Range
0 ... 999999 s
Default value
1 s
Start volume
Start volume
Volume to be added before the start of the titration at the dosing rate indi-
cated.
Titrando, 855
Range
0.00000 ... 9999.99 mL
Default value
0.00000 mL