beautypg.com

Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 534

background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

518

■■■■■■■■

tiamo 2.1 (ProcessLab)

Titrando, 855
Range

0.01 ... 99.90 µA/min

Default value

20.00 (slow), 50.00 (optimal), 80.00 (fast) µA/
min

Selection

off

off
Measured value acceptance will take place after the maximum waiting
time has elapsed.

Titrino
Range

0.05 ... 99.90 µA/min

Default value

20.00 (slow), 50.00 (optimal), 80.00 (fast) µA/
min

Selection

off

off
Measured value acceptance will take place after the maximum waiting
time has elapsed.

Min. waiting time

The minimum waiting period is only important for drift-controlled measure-
ments. The measured value is not accepted until the minimum waiting time
has elapsed, even if the signal drift has already been reached. The drift con-
tinues to be checked while the waiting time is elapsing.

Range

0 ... 999999 s

Default value

0 s

Max. waiting time

If signal drift has been switched off or has not yet been reached, then the
measured value will be accepted when the maximum waiting time has
elapsed.

Range

0 ... 999999 s

Default value

1 s

Start volume

Start volume

Volume to be added before the start of the titration at the dosing rate indi-
cated.

Titrando, 855
Range

0.00000 ... 9999.99 mL

Default value

0.00000 mL