5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
942
■■■■■■■■
tiamo 2.1 (ProcessLab)
5.6.4.9.3
MEAS Cond - Measuring parameters
Tab: Method
▶ MEAS Cond ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Input
25 characters
The parameters for the measuring procedure can be defined on this tab.
Measurement
Measurement frequency
712
Selection
auto | 300 Hz | 2.4 kHz
Default value
auto
auto
The more suitable of the two frequencies is selected.
Type of measurement
The two measuring modes Measurement with drift control or Meas-
urement without drift control can be selected.
Selection
Measurement with drift control | Measurement
without drift control
Default value
Measurement with drift control
Measurement with drift control
The measurement is carried out drift-controlled if this option is enabled.
The measurement is stopped as soon as the defined Signal drift or the
Stop measured value is reached or a waiting time has elapsed. The
corresponding parameters are not visible if this option is disabled. This
option is not available for the 712 Conductometer.
Measurement without drift control
The measurement is carried out without drift control if this option is ena-
bled. The measurement continues until one of the stop criterions
Measuring time or Stop measured value is met. The corresponding
parameters are not visible if this option is disabled.