5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

989

Note

If an existing solution is selected as the solution to be added then the
Concentration and Unit of this solution will be adopted for calculating
the standard addition. If not defined is selected, then Concentration
and Unit must be entered on the Standard addition tab.

Sensor

Measuring input

Selection of the measuring input to which the sensor is connected.

Titrando, 867
Selection

1 | 2

Default value

1

855
Selection

1

Default value

1

Sensor

Selection of a sensor of type ISE electrode from the sensors configured in
the sensor table.

Selection

'Sensor name' | ISE electrode

Default value

ISE electrode

Temperature measurement

Type of temperature measurement.

Selection

continuous | automatic | off

Default value

automatic

continuous
A temperature sensor must be connected. The temperature will then be
measured continuously.
automatic
If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise, the temperature entered manually under
Temperature on the Measuring parameters tab will be used.
off
The temperature will not be measured. The temperature manually
entered under Temperature on the Measuring parameters tab will
be used.