2 cal meas conc - general/hardware – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 1036
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1020
■■■■■■■■
tiamo 2.1 (ProcessLab)
Identifica-
tion
Description
.ITE
Initial temperature (temperature before start conditions
are processed) in °C
.LP.MEA
Measured value for the last measuring point in the meas-
uring point list in the unit of the measured value
.LP.TEM
Temperature for the last measuring point of the measur-
ing point list in °C
.LP.TIM
Time in s until the last measuring point in the measuring
point list is reached
.MTE
Temperature measurement with sensor; 1 = on, 0 =
off)
.NMP
Number of measuring points in the measuring point list
.STY
Type of stop with which the command was stopped: 1 =
normal;0 = manual or after error
5.6.5.7.2
CAL MEAS Conc - General/Hardware
Tab: Method
▶ CAL MEAS Conc ▶ Properties... ▶ General/Hardware
Command name
Name of the command.
Input
25 characters
The general parameters for the device, the sensor and the stirrer are defined
on this tab.
Device
Device name
Selection of a device from those available in the device table. The only devi-
ces which shall be offered are those with which the command can run.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start of
the method.
Device type
Display or selection of the device type. If a device is selected under Device
name, then this Device type field can no longer be edited, and the device
type belonging to the selected device is displayed instead