8.4 Manual Control - Functions
■■■■■■■■■■■■■■■■■■■■■■
1470
■■■■■■■■
tiamo 2.1 (ProcessLab)
8.4.1.5
Dosing - Empty
Program part: Manual Control
Here the emptying of dosing units can be started and stopped. This tab
appears when selecting a single dosing device as well as with all dosing
devices with dosing units.
Note
This tab is not displayed for external dosing devices on the Titrinos 736,
751, 758, 799.
[Start]
Start emptying for the selected dosing device(s). The parameters defined for
the dosing units will be used.
[Stop]
Stop emptying for the selected dosing device(s).
8.4.1.6
Dosing - Add fix volume
Program part: Manual Control
Here the dosing of a preset volume can be started and stopped. This tab
only appears if a single dosing device is selected.
Volume
Fixed volume to be dosed.
Range
100 ... 99999.9 ml
Default value
100 ml
Dosing rate
The volume is dosed at this rate. The maximum dosing rate depends on the
cylinder volume of the exchange unit or dosing unit used. If the entered
dosing rate is too high for the selected dosing device, it will automatically
be reduced during dosing to the largest possible value.
Note
For viscous liquids the dosing rate should be reduced.
Selection
maximum
Default value
maximum
Titrando, 814, 815, 855
Range
0.01 ... 166.00 mL/min