8 stat ph - evaluations, Evaluations – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 842
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
826
■■■■■■■■
tiamo 2.1 (ProcessLab)
[Properties]
Opens the dialog window Track call # in which the parameters for the
track call can be edited (see Chapter 5.6.3.8.4, page 802).
[Delete]
Deletes the track call selected in the table.
5.6.3.8.6.8
STAT pH - Evaluations
Tab: Method
▶ STAT pH ▶ Properties... ▶ Evaluations
Command name
Name of the command.
Input
25 characters
The following methods for the evaluation of titration curves can be activated
and defined on this tab:
Rate evaluation
Rate evaluation
on | off (Default value: off)
If this check box is activated, than a maximum of 9 time slots can be defined
in which the mean dosing rate can be determined by linear regression. The
defined time slots are shown in the window table and can be edited with
the following buttons:
[New]
Opens the dialog window Evaluation window # in which the parameters
for a new time slot can be entered (see Chapter 5.6.3.8.3, page 802).
[Properties]
Opens the dialog window Evaluation window # in which the parameters
for the new time slot can be edited (see Chapter 5.6.3.8.3, page 802).
[Delete]
Deletes the time slot selected in the table.
Fix endpoint evaluation
Fix endpoint evaluation
on | off (Default value: off)
If this option is enabled, then the associated values for the other measured
quantities will be interpolated with a fixed measured quantity (measured
value, time or volume) for the fix endpoint from the measuring point list.
Moves the selected line up (changes the sequence).