Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 1300
6.7 Subwindow Sensors
■■■■■■■■■■■■■■■■■■■■■■
1284
■■■■■■■■
tiamo 2.1 (ProcessLab)
Sensor monitoring
on | off (Default value: off)
If this option is switched on then the working life of the sensor will be moni-
tored.
Working life
Working life of the sensor in days. If a value is entered here then the Expiry
date will be automatically adapted.
Range
0 ... 999 Days
Default value
999 Days
Expiry date
Expiry date of the sensor. The date can be selected by pressing
in the
dialog window Select date (see Chapter 2.5.1, page 79). When a date has
been entered the Working life will be automatically adapted.
Can only be edited if monitoring is switched on.
Set value
Set to work+ 999 days
Message
The message options can only be edited if monitoring is switched on.
Message by E-mail
on | off (Default value: off)
The message is additionally sent to the E-mail address defined under [E-
mail] if this option is activated. The message is sent in text format.
[E-mail]
Pressing [E-mail] opens the Send E-mail (see Chapter 2.6.1, page 81)
window.
Acoustic signal
on | off (Default value: off)
If this option is enabled, an acoustic signal will be emitted additionally to the
message defined above
Action
The actions can only be edited if monitoring is switched on.
If during sensor monitoring it is found that the working life has expired, then
one of the following actions will be triggered automatically at Start test: