9 det upol - measured value window – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 549
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.1 (ProcessLab)
■■■■■■■■
533
can also be displayed in curves under the designation Extern 1…3 and used
in formulas as variables 'Command name.EX1…3'-
[New]
Opens the dialog window External measured value # in which the
parameters for the external measured value can be entered (see Chapter
5.6.3.10.2, page 859).
[Properties]
Opens the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.10.2, page 859).
[Delete]
Deletes the external measured value selected in the table.
5.6.3.2.5.9
DET Upol - Measured value window
Dialog window: Method
▶ DET Upol ▶ Properties... ▶ Potentiometric evalua-
tion
▶ Evaluation with measured value window (Upol) ▶ [New]/[Proper-
ties]
▶ Measured value window#
Measured value windows are regions on the measured value axis for which
different parameters for potentiometric evaluation can be defined. Only
those endpoints which lie within these windows and additionally meet the
parameters defined for each window will be recognized. Only one endpoint
will be recognized per window.
Lower limit
Lower limit for the measured value window.
Range
-200.0 ... 200.0 µA
Default value
-200.0 µA
Upper limit
Upper limit for the measured value window.
Range
-200.0 ... 200.0 µA
Default value
200.0 µA
EP criterion
Criterion for the recognition of equivalence points. Equivalence points
whose ERC is smaller than the set EP criterion will not be recognized.
Range
0 ... 200
Default value
5
EP recognition
Filter for the recognition of equivalence points: